Please see also TofPCBComments27Nov2019 as many things there are not done yet.
HV Board
- Change R20 and R21 to 10M and 1M respectively. Also change them to a larger footprint or thru-hole which is rated for > 200V
- Ground wiring around U8 and U10 is unacceptable. Use one GND via minimum for each 1-2 connections.
- See http://ohm.bu.edu/~hazen/elbat_tof/2019/HV_gnd_layout.pdf
- Similar story for -5V, -195V and
- Same for U9 and U11 but use -195V instead