LT584 programming notes
LT584 waveform template: template.txt read from instrument with "TEMPLATE?" command.
'Initialization
Initialization sequence extracted from VB software (MainDAQ1.zip
"Write" means to send a command and not expect a response.
"Query" means to send a command and read back a response.
- Write Device Clear (++clr), wait 1 seconds
- Write "*RST"
- Wait
- Query "*IDN?" and read identification string
- Write "WFSU SP,0,NP,500000,FP,0,SN,0" (Waveform setup: send all data points, all segments)
- Write "TRSE EDGE,SR,C1" (Trigger setup: edge trigger, channel 1
- Wait
- Write "TRDL 20PCT" (trigger delay: 20 percent)
- Write "C1:TRCP AC" (trigger coupling)
- Write "C1:TRLV 0.01V" (trigger level)
- Write "C1:TRSL POS" (trigger slope)
- Wait
- Write "TRMD SINGLE" (single sweep mode)
- Wait
- Write "TDIV 1.0US" (horizontal scale)
- Wait
- Write "C1:TRA ON" (trace on)
- Write "C1:VDIV 0.1V" (vertical scale)
- Write "C1:ATTN 1" (1x attenuation)
- Write "C1:CPL A1M" (AC 1M coupling)
- Write "C1:OFST 0V" (no DC offset)
- Wait
- Write "SEQ ON,1000,500" (1000 segments of 500 points)
- (delay 500ms)
- Wait
The "Wait" operation can consist of sending the "*OPC?" command and waiting for the response (which should always be "1").
'Arm for Trigger
- Write Device Clear (++clr)
- Write "ARM"
'Check for Trigger
- Query "INR?" (wait for bit 13 8192?
'Read trigger times
- Query "C1:INSP? TRIGTIME'". Returns list of times (try it to get the format).