Changes between Initial Version and Version 1 of KFLARE_Operating_Notes
- Timestamp:
- Nov 8, 2013, 10:17:54 AM (12 years ago)
Legend:
- Unmodified
- Added
- Removed
- Modified
-
KFLARE_Operating_Notes
v1 v1 1 Back to [[KFLARE]] 2 3 === Laser Current Limit Setting === 4 5 Attach a voltmeter to J12 (for channel 0) and adjust R40. 6 The scale is 0.5V/A so to set to a limit of 2200mA for example, set to 2.2*0.5 = 1.1V on the test point. 7 8 For channel 1, measure on J16 and adjust R73. 9 10